What it measures
Equal parts of oil and aniline are mixed; at room temperature they form two phases. On heating, they reach a point where they become a single transparent phase. That temperature, or more precisely the temperature at which turbidity appears on cooling, is the aniline point.
How it is interpreted
Aniline is an aromatic solvent. An oil rich in aromatic compounds mixes with it easily, and therefore at a lower temperature. So:
- Low aniline point: more aromatics, greater solvent power.
- High aniline point: fewer aromatics, more saturated paraffinic or naphthenic character, lower solvent power.
Why it matters in electrical equipment
- Material compatibility: an oil with greater solvent power can swell or soften gaskets and elastomers, and can affect internal varnishes or paints. It is a criterion when replacing or mixing oils.
- Dissolution of aging products: solvent power influences how much sludge and how many polar compounds the oil keeps in solution.
- Stability and gassing: aromatics affect oxidation resistance and the behavior of the oil under electrical discharges, so aromatic content is a formulation compromise rather than an absolute virtue or defect.
What an engineer sees in practice
The parameter appears on the product data sheet or when comparing two candidate oils, for example when topping up equipment with an oil from another source. It is rarely used to monitor an in-service oil, because its variation with aging is small compared with acidity or interfacial tension.
A laboratory safety consideration
Aniline is a toxic substance; the test must be performed with adequate ventilation and protection in accordance with the safety data sheets. Acceptance values depend on the specification; consult the current standard.