What it is
The term applies mainly to the portable dielectric strength tester. It is a compact instrument with a step-up transformer, a cell with standardized electrodes, a control system that raises the voltage uniformly, and a circuit that detects the instant of the arc and cuts off the supply.
How it works
- The cell is filled with bubble-free oil, making sure the vessel and electrodes are clean.
- It is left to stand for a time so the bubbles can escape.
- The instrument raises the voltage at a fixed rate until breakdown occurs and records the value.
- After each breakdown, the oil is stirred, the operator waits, and the test is repeated until the number of determinations required by the method is complete; the result is the average.
What an engineer sees in practice
A value in kilovolts for each breakdown and an average. A low result points to moisture, particles, or contamination in the oil. A high result means the oil is adequate in that respect, but nothing can be concluded about acidity, gases, or the condition of the paper.
Where it fits among the tests
It is a quick screening test that helps decide whether to filter or dry on site, or whether to send a sample to the laboratory. It does not replace full analysis and may differ from the laboratory value because of handling, temperature, or the cell method.
Limitations
- It is very sensitive to the way the cell is filled and to bubbles.
- The cell, the electrodes, and the gap between them must match the chosen method.
- It requires periodic verification.
Safety
The tester generates high voltage and works with oil that can ignite after repeated arcs. It must be operated by qualified, trained personnel, following the manufacturer’s instructions and the applicable regulations.