Dielectric oil test equipment
We rent equipment for oil tests such as dielectric breakdown, density, moisture, power factor or tan delta, and dissipation factor. The equipment is in stock and immediately available at our laboratory in Monterrey, Mexico.
Standard and user-defined tests are fully automated and cover standards such as ASTM D877, D1816, D1533, D4052, D971, D1524, D445, D1169, D974, D924, D7042.
Equipment features
- Integrated heating and cooling systems for the test cells.
- Efficient oil filling and integrated automated oil draining.
- IEC 60247 test cell: easy to assemble and clean.
- Stores up to 50 records on board and downloads to a PC.
- Compact and lightweight (32 kg.).
Oil Tan Delta (OTD)
The Oil Tan Delta (OTD) is a laboratory instrument that measures the Tan Delta, resistivity and relative permittivity of insulating liquids. These tests give a broader picture of the overall quality of the insulating oil and indicate whether contamination is present. Megger developed the OTD’s features to increase productivity in laboratories.
The microprocessor-controlled induction heater maintains temperature uniformity throughout the cell, in accordance with IEC 60247. For cooling, a fan speeds up the cell’s cooling time.
The instrument measures temperature from 10 to 110 ºC, with an accuracy of ± 1 ºC. It can be set to preheat before the test. Automated draining eliminates the need to move the oil container.
Safety and cell design
The equipment has independent, doubly redundant high-voltage cut-off circuits. If the user needs to abort the test, this can be done by pressing any key: the test ends immediately and the high voltage is removed.
The design of the cell components maximizes the oil filling rate, minimizes flow turbulence and eliminates air build-up. With only 32 cell components, it is easy to assemble and disassemble.
Calibration verification
With the calibration verifier, you can perform a quick check to validate the Tan Delta, resistivity and permittivity measurements. An on-site service from an authorized calibration service center makes it possible to adjust the measurements when necessary and minimize downtime.