Basic idea
Gas chromatography exploits the fact that each gas interacts differently with the material in a column. An inert carrier gas, such as helium, nitrogen, or argon, sweeps the mixture through the column. The components that interact least come out first and those that interact most come out later. That difference in times separates the mixture.
Parts of an instrument for oil
- Extraction system: releases the dissolved gases from the oil so they can be injected.
- Injection valves: introduce an exact volume of gas into the carrier flow.
- Columns: packed or capillary, with different materials to separate the permanent gases and the hydrocarbons.
- Oven: keeps the column temperature stable.
- Detectors: the thermal conductivity detector detects hydrogen, oxygen, and nitrogen; the flame ionization detector detects hydrocarbons; a methanizer converts carbon monoxide and carbon dioxide into methane so that the flame detector can see them.
What an engineer sees in practice
The instrument produces a chromatogram, a plot of peaks over time. The position of each peak (retention time) identifies the gas, and its area, or height, is proportional to its quantity. The software calculates the concentration by comparing it with the response to a standard mixture of certified composition.
What keeps the measurement reliable
- Calibration with standard gases at defined intervals.
- Carrier gas of adequate purity, free of leaks and moisture.
- Verification of resolution between closely spaced peaks.
- Temperature control of columns and detectors.
How to interpret the result
A peak that is shifted, broadened, or tailing warns of an aged column or a leak, and must not be quantified until corrected. The results feed into dissolved gas diagnosis.