What is determined
The neutralization number, or acid number, quantifies the acidic compounds in the oil. It is defined as the mass of potassium hydroxide needed to neutralize the acids present in one gram of sample, and it is reported in milligrams of KOH per gram.
Measurement principle
It is an acid-base titration. A weighed quantity of oil is dissolved in a solvent mixture that allows the oil and the alkaline solution to come into contact, for example a mixture of toluene, isopropyl alcohol and a little water. A standard potassium hydroxide solution of known concentration is added little by little until the acids are neutralized.
How the end point is detected
- Colorimetric: an indicator changes color on reaching neutrality. It works well with light-colored oils, but becomes difficult with very dark oils.
- Potentiometric: an electrode measures the potential of the solution, and the end point is identified by the abrupt change in potential. It is more objective and works with dark samples.
What an engineer sees in practice
The laboratory reports a numerical value that is compared with the new oil and with previous results. The values for insulating oils are small, so accuracy depends on weighing the sample correctly, having fresh, standardized reagents, and running a solvent blank.
Sources of error
- A sample that is not representative or contains water and sediment.
- Degraded or poorly standardized alkaline reagent.
- Container contamination with acidic or alkaline residues.
- Subjective visual judgment of the color change.
How to interpret the result
What matters is the trend relative to the equipment’s history, together with interfacial tension and color. The titration measures total acidity, without distinguishing between strong and weak acids.