How Is a Conformal Coating Evaluated?

A conformal coating is evaluated by confirming that it covers continuously and uniformly, that it adheres, that it maintains its insulation under moisture and temperature, and that it protects against contamination and corrosion without damaging the circuit.

What is expected of the coating

It is a thin film on an electronic circuit board that protects it from moisture, dust, contamination and corrosion, and that increases the insulation between closely spaced traces. Evaluating it means verifying that it performs that function continuously and durably.

Coverage and thickness inspection

It is inspected visually, often under ultraviolet light if the coating contains a fluorescent tracer, to detect uncovered areas, bubbles, orange peel and pooling. Thickness is measured with methods suited to a film on a substrate, or on cross sections. Corners, component edges and leads are the critical points, because surface tension pulls the coating away from sharp edges.

Adhesion and compatibility

Adhesion depends on prior cleaning: flux residues or oils create poorly adhered areas. It is verified with adhesion tests, and compatibility with the board, components and fluxes is examined. The solvents in the coating can affect certain plastics.

Electrical and environmental performance

  • Surface insulation resistance between traces, before and after moisture exposure.
  • Thermal cycling: looks for cracks or lifting caused by the difference in expansion with components and board.
  • Moisture and condensation: with electrical bias to detect corrosion and electrochemical migration.
  • Salt fog or corrosive gases, depending on the environment of use.

An example of interpretation

If the insulation resistance falls after moisture exposure, first check for uncovered areas or contamination trapped under the film: a coating over ionic residues can even aggravate the problem by retaining moisture.

Other aspects

  1. Effect on high-frequency signals, because the film adds capacitance and losses.
  2. Heat dissipation and stresses on components.
  3. Reparability and removal of the film.
  4. Areas that must remain uncoated, such as connectors and test points.

General criterion

The evaluation is defined according to the applicable product standard and the actual conditions of use.