What does IEC 60093 define for resistivity?

IEC 60093 establishes methods for measuring the volume and surface resistivity of solid electrical insulating materials, defining electrodes, DC voltage, electrification time, conditioning and calculation.

General scope

The standard deals with measuring the volume resistivity and surface resistivity of solid insulating materials, in the form of sheets, plates or flat parts, using DC voltage. It defines how the sample is placed, which electrodes are used and how the measured resistance is converted into resistivity.

What it contains

  • Definitions: volume resistivity, associated with current passing through the material, and surface resistivity, associated with current passing along the surface. They are distinct quantities that must not be mixed up.
  • Electrode arrangement: measuring electrode, guard electrode and counter electrode, which separate the volume current from the surface current. The choice determines how the effective area is calculated.
  • Voltage and time: DC voltage is applied and an electrification time is allowed before reading, because the current decays at first due to polarization.
  • Conditioning and ambient conditions: humidity and temperature have a strong influence, particularly on surface resistivity.
  • Calculation and report: how to calculate and which data must be reported.

What to bear in mind

Surface resistivity is not a property of the material alone: it depends on the cleanliness, the humidity and the condition of the surface. That is why the value is highly variable and is interpreted in the context of the conditioning. Volume resistivity is more stable, although it also depends on temperature and voltage.

Relationship to other standards

There are standards from other bodies with the same objective and differences of detail in electrodes and times. Results from one and the other may not be equivalent, so they must not be compared without verifying the conditions.

How to use it

  1. Consult the current edition before defining the procedure.
  2. Report the voltage, electrification time, electrodes and environment.
  3. Use an instrument with adequate resolution for very small currents.