What is measured
The rate of change is the temperature variation per unit of time during a heating or cooling ramp. It matters in thermal shock and cycling tests, because the fatigue a sample undergoes depends in part on how fast its temperature changes.
Chamber air versus the sample
The chamber may achieve a fast ramp in the air, but the sample, with its thermal mass, follows with a lag. A heavy part may change slowly even when the chamber meets the programmed ramp. It must therefore be made clear whether the specified rate refers to the air or to the sample, and the measurement taken in the right place.
How it is measured
- Place sensors in the air of the working zone, in several positions, and on the sample or on a representative mass.
- Run the ramp with the defined load: empty or with the test mass.
- Record at an interval that is short relative to the ramp duration.
- Calculate the slope over the section the procedure defines, either as an average of the entire ramp or as a mean value in a moving window.
- Verify that the result meets the required rate and tolerance.
What influences it
- The thermal load: the greater the mass, the slower the effective ramp.
- Air circulation: a well-distributed flow reduces differences.
- The temperature range: cooling or heating capacity decreases near the extremes.
- The sample’s own heat source, if it is energized.
What the engineer sees
The real curve has an initial start-up section, a nearly linear zone and a deceleration on approaching the setpoint. Calculating the slope over the entire ramp gives a lower value than that of the linear zone, so it is advisable to set in advance how it will be calculated. Results from different chambers are comparable only if the same calculation and load criteria are used.