How temperature distribution in a chamber is validated

A grid of calibrated sensors is placed in the working volume, the chamber is stabilized, and the deviation and variation of each point from the setpoint are analyzed over a period long enough to reveal drift.

What needs to be known

A chamber can be accurate at the control sensor and still have colder or hotter zones. Distribution validation quantifies two things: spatial uniformity, that is, how much the points differ from one another at a given moment, and temporal stability, how much each point fluctuates over time.

How it is carried out

  1. Plan: define the working volume, the test temperatures and the load condition (empty or with a representative load).
  2. Distribute the sensors: place them in the corners, at the center and near the control sensor location, with a density suited to the size of the chamber. Corners and areas near doors are usually the worst cases.
  3. Calibrate the sensors beforehand and, if the procedure requires it, check them afterward.
  4. Stabilize the chamber until the control settles.
  5. Record for a period that covers several control cycles.
  6. Analyze: the difference between maximum and minimum across sensors, the deviation of each one from the setpoint, and the variation over time.

How to interpret it

  • If all sensors deviate equally from the setpoint, there is a control bias.
  • If some points depart from others, there is a gradient caused by air circulation or design.
  • If the signal oscillates periodically, that is the control cycle.

Factors that alter the result

The load blocks airflow and changes the profile. Opening the door disturbs the state. The thermal mass of the load delays stabilization. For these reasons the validation must be done under conditions that represent actual use, and repeated after repairs or major changes.

How to use the result

The acceptance criterion depends on the standard or the process requirement, and the values must be taken from that document, not assumed. The result also makes it possible to locate the zones where critical samples are best placed and to report the associated uncertainty.