Short answer
IEC 60584 covers the standardization of thermocouples: for each letter-designated type it defines how much voltage the thermocouple must generate at each temperature, relative to a reference junction, and how far a real thermocouple may deviate from that relationship.
Why it is needed
A thermocouple produces a small voltage through the Seebeck effect, which depends on the temperature difference between the measuring junction and the reference junction. For thermocouples from different manufacturers and a given readout instrument to be compatible, the relationship between that voltage and temperature has to be fixed. The standard does this through reference functions and tables for each type.
General content
- Reference functions and tables relating electromotive force and temperature for each type.
- Tolerance classes limiting how far a new thermocouple may deviate from those tables.
- Identification of the types by letter, and their alloy composition.
Tolerances depend on the current edition and on the type, so they should be looked up in the document rather than quoted from memory.
What the standard does not resolve
The table describes an ideal thermocouple. In practice, the actual error depends on the homogeneity of the wire, which changes with use at high temperature, on the gradient along the cable, on the reference junction compensation and on the extension cable. Two thermocouples of the same type that both meet the class can differ from each other in the process.
What an engineer sees
The readout instrument applies the function from the standard to convert millivolts to temperature. If the wrong type is configured, the reading comes out with a large, systematic error. If an extension cable or connector of another type is used, a parasitic junction appears and alters the signal.
How to use it
Select the type according to range, atmosphere and stability, confirm the required class in the current edition, and verify that the readout, cable and connectors all match the same type. Calibration remains the way to find out the error of an installed thermocouple.