What it is for
Calibrating a temperature sensor by comparison requires a medium in which everything immersed is at the same temperature and that temperature remains stable. A fluid calibration bath provides that medium.
How it works
It consists of a tank with a liquid, a heating system and, depending on the model, a cooling system, a temperature controller and a stirrer. The liquid circulates, so that heat is distributed and the temperature differences between points in the tank are small. The sensors are immersed to the necessary depth, next to the reference thermometer, and their readings are compared once the temperature has stabilized.
Why a liquid
A stirred liquid transfers heat to the sensor very well, more uniformly than air or a block with clearances. It accepts sensors of different shapes and sizes without the need for special inserts, and the sensor’s response is usually fast, which shortens the calibration.
Fluids by range
- Water: for moderate temperatures.
- Silicone or similar oils: for higher temperatures, with ranges of use defined by the oil manufacturer.
- Mixtures with alcohol or other low-temperature fluids: for temperatures below ambient.
- Molten salts: for very high temperatures, with special precautions.
The fluid must be chosen so that its range of use, viscosity, flash point and compatibility with the sensors are suitable.
Qualities evaluated
- Stability: how much the temperature varies over time at one point.
- Uniformity: temperature differences between points in the working volume.
- Working depth: the region where uniformity is adequate.
When it does not apply
When the sensor cannot get wet, when a portable instrument is required for field use, or when spills and vapors are unacceptable, a dry-block calibrator is preferable. In exchange, the block usually offers lower uniformity.
Maintenance and safety
The fluid ages, becomes contaminated and absorbs moisture, and must be replaced according to the manufacturer’s instructions. At high temperatures there is a risk of burns, vapors and fire, and a wet sensor must not be placed in hot oil because the boiling water can spatter. Operation must be carried out by trained personnel.