Why it is a delicate measurement
Relative humidity is the ratio between the partial pressure of water vapor in the air and the saturation vapor pressure at the same temperature. Because saturation pressure rises rapidly as temperature increases, a small temperature error becomes a significant relative humidity error. That is why accuracy in humidity is, to a large extent, accuracy in temperature.
Conditions for a good measurement
- Thermal equilibrium. The sensor and the air it measures must be at the same temperature. A sensor just brought from another environment, or heated by its own electronics, will give a wrong value. It is advisable to wait for it to stabilize.
- Good sampling. The probe must be in the air of interest, with enough circulation, away from heat sources, direct radiation, cold walls and air outlets that do not represent the zone.
- A suitable sensor. Capacitive polymer sensors are the most common: the water content of the polymer changes its capacitance. Their performance depends on the quality of the polymer and on temperature compensation.
- Sensor protection. Filters keep out dust and splashes, but they slow the response.
- Calibration conditions. Calibration is done at known points, with saturated salt solutions in closed chambers or with a humidity generator, and verified with a reference standard such as a chilled-mirror hygrometer.
Phenomena that degrade the measurement
- Hysteresis. The sensor reads differently depending on whether it comes from dry or humid environments.
- Contamination. Chemical vapors, oils and particles alter the response.
- Condensation. A wet sensor needs time to recover, and may remain offset.
- Slow response at low temperatures. Vapor equilibrium takes longer.
Common errors
- Measuring close to the operator’s body or the hand holding the instrument.
- Comparing two hygrometers that are not at the same point or at the same temperature.
- Reading immediately after a change of environment.
- Demanding more accuracy than the sensor can offer at the extremes of its range.
How to interpret the result
A small difference between two hygrometers can come entirely from a temperature difference of tenths of a degree. To decide, compare the temperature of both and calculate what relative humidity that difference would imply before concluding that either one is off.