How is a resistance thermometer calibrated by comparison?

It is calibrated by comparing the resistance thermometer's readings with those of a reference standard, both immersed in a stable isothermal medium, at several temperatures spread across the range of use.

Principle

Calibration by comparison does not measure temperature in an absolute way. The thermometer under calibration and a reference standard are placed in a medium where both are considered to be at the same temperature, and their indications are compared. The sensor’s error is the difference from what the standard indicates, and the standard is itself traceable.

What you need

  • An isothermal medium. A stirred bath or a dry-block calibrator with a suitable insert, whose temperature is uniform and stable in the measurement zone.
  • A reference thermometer. Calibrated, with an uncertainty clearly smaller than the sensor’s tolerance.
  • A resistance readout instrument, of sufficient accuracy, ideally measuring both sensors with the same technique.

General steps

  1. Inspect the sensor and its wiring and record its identification.
  2. Place the standard and the sensor side by side, at the same depth, with enough immersion that heat conduction along the stem does not distort the measurement.
  3. Set the first temperature and wait for the medium and both sensors to stabilize.
  4. Take several readings of both, preferably alternating, to detect drift in the medium.
  5. Repeat at each calibration temperature.
  6. Calculate the error at each point and estimate the uncertainty.

Sources of error to control

  • Insufficient immersion. Part of the heat is lost along the stem, and the sensor does not reflect the temperature of the medium.
  • Gradients in the medium. If the block or bath is not uniform, the standard and the sensor do not see the same temperature.
  • Self-heating. The measuring current heats the element; it must be the same for both and appropriate.
  • Lead resistance. It is compensated with a four-wire or three-wire connection, as the case requires.
  • Incomplete stabilization. It is advisable to know the sensor’s time constant.

How to interpret the result

The pairs of readings give the error at each point, which is compared with the tolerance, taking into account the uncertainty of the standard, the medium and the readout instrument. If the sensor serves at a single process point, the calibration can be limited to the area around it; if it is used over a range, the extremes and intermediate points must be covered.

Heating equipment and high-temperature media involve a risk of burns; the work must be done by trained personnel.