What Is a Voltage Reference Source?

A voltage reference source is a device that delivers a very stable, known voltage, used as a standard or as a point of comparison to calibrate voltmeters, converters and other instruments.

Its function is to supply a voltage value that changes as little as possible with time, temperature and load. With that known value, other instruments are calibrated or a laboratory’s electrical scale is maintained between external calibrations. It is not the same as a power supply: a power supply delivers power with acceptable regulation; a reference delivers a precise value at very low current.

How it achieves its stability

Most electronic references are based on a special Zener diode or on bandgap reference circuits, with temperature compensation. For better stability they are held at a controlled temperature inside an enclosure, which reduces the effect of the environment. Even so, they drift slowly over time, so they are calibrated periodically and their history is tracked.

Characteristics that matter

  • Long-term stability: the drift between calibrations.
  • Temperature coefficient: how much it changes with the environment.
  • Noise: rapid fluctuations of the value.
  • Load regulation: how it varies when an instrument with low input impedance is connected.
  • Warm-up time: many need to be on for some time before reaching their stability, and some are kept powered at all times.

What the engineer sees in practice

It is used to verify a voltmeter: the reference is connected and the reading is compared. In laboratories with a group of references, they are compared with one another to detect whether any one drifts more than the others. Transport and sudden temperature changes can temporarily alter the value, so the reference is left to rest before use.

What to check first

Check that the certified value corresponds to the conditions of use, that the connection has clean contacts and suitable leads, and that the calibration date is current. Work with high voltages must be carried out by qualified personnel in accordance with the applicable electrical safety regulations.