An ordinary multimeter is used for diagnostics: it shows whether voltage is present or whether a circuit conducts. A reference multimeter works on the same measuring principle but with higher resolution, better stability and lower uncertainty, and it is kept under traceable calibration. It is used for comparing and calibrating, not for field work.
What sets it apart
- High resolution: it shows more useful digits, which makes small differences visible.
- Stability: its response changes very little with time and temperature, which makes its certificate reliable between calibrations.
- Low nonlinearity and noise: better than in general-purpose equipment.
- Input characteristics: high input impedance on DC voltage so as not to load the circuit being measured.
- Additional functions: four-wire resistance measurement, which eliminates the effect of the leads.
Common uses
Verifying the output of voltage and current sources, calibrating lower-class multimeters, checking multifunction calibrators, measuring reference voltages, or evaluating resistors. In each case its function is to be the standard against which the comparison is made.
What the engineer sees in practice
The quality of the result depends on the setup: suitable leads, clean connections, no thermal voltages at the copper contacts and warm-up time before measuring. An excellent multimeter with poor technique produces deficient measurements. The appropriate range must also be used, because uncertainty is specified by range and function.
Limitations
- It does not replace a reference source when a value must be generated; it only measures.
- Its uncertainty is useful only if its calibration is current and covers the functions used.
- For very special quantities, such as very low voltages or very high resistances, specific instruments may be required.
When working with hazardous voltages, the work must be done by qualified personnel following the applicable electrical safety regulations.