Linearity is determined by measuring the instrument's response at several points across the span, fitting a reference line to the data and calculating the maximum deviation of the points from that line.

An instrument is linear if its output is proportional to its input across the whole span. Real linearity is never perfect, and it is quantified as the maximum deviation between the measured response curve and a reference line. Everything depends on which line is chosen as the reference, and that choice must be stated.

Procedure

  1. Using a traceable standard, apply values distributed across the span, with enough points to see the shape of the curve and not just the ends.
  2. Record the instrument output at each point, preferably upscale and downscale, and with repeats.
  3. Choose the reference line.
  4. Calculate the deviation of each point from the line.
  5. Express the largest deviation, in absolute value, relative to the span or to full-scale output.

Types of reference line

  • End-point (terminal) line: joins the lowest and highest points. It is simple, but usually gives larger deviations.
  • Least-squares line: minimizes the sum of the squares of the deviations and is the most widely used in data analysis.
  • Best-fit line in terms of maximum deviations: distributes the positive and negative deviations so that the largest is as small as possible.

The same curve gives different linearity values depending on the line, so comparing manufacturers’ specifications requires knowing which one was used.

What the engineer sees in practice

If an instrument’s error at several points follows a smooth curve that moves away from the line in the middle of the span, zero and gain adjusted at the ends do not correct the middle. That is a lack of linearity, and basic adjustment does not resolve it. Some solutions are a correction by table or polynomial, or calibrating only over the range of use.

When it does not apply

Some sensors are nonlinear by design, such as certain thermistors or differential-pressure flow sensors. For these, the complete curve is characterized and an equation or table is used; linearity is not treated as a defect.