Hysteresis is the dependence of the indication on the previous path: the same input value produces different readings depending on whether it is reached going up or going down. It is measured with a complete cycle across the span, comparing both directions point by point.
Procedure
- Stabilize the instrument and check the zero.
- Go up through the chosen points from lowest to highest, reading at each one after stabilization.
- Reach the maximum value of the cycle and wait.
- Go down through the same points, from highest to lowest, reading again.
- At each point, calculate the difference between the downscale and upscale readings.
Critical precautions
- Do not overshoot: if you overshoot the value on reaching a point and back up, you have mixed both directions and the point is unusable; the cycle must be repeated.
- Same reference value: the standard must reproduce the points with sufficient stability.
- Waiting time: in instruments with a slow response or creep effects, measuring too soon confuses hysteresis with drift over time.
- Tapping: some mechanical pressure gauges are tapped gently before reading; the procedure must be the same at all points.
Why it happens
In mechanical instruments, because of friction and imperfect elasticity of the sensing elements. In pressure and force sensors, because of the behavior of the diaphragm or load cell material. In magnetic materials, because of magnetization itself. It is an effect inherent to the design rather than an accidental defect, although its magnitude varies with the quality and condition of the instrument.
How to interpret the result
Hysteresis cannot be eliminated by adjusting zero or gain, because it is not a constant offset. If it is large compared with the tolerance, the instrument must always be used in the same direction of approach, or be replaced. It also forms part of the uncertainty when the direction of measurement is not controlled in use.