There is no single correct interval valid for all instruments of the same type. The interval is a decision that balances two risks: calibrating too often costs time and money, and calibrating too rarely increases the likelihood of unknowingly using an out-of-tolerance instrument. The starting point is usually the manufacturer’s recommendation, which is then refined with your own evidence.
Influencing factors
- Stability: how much the instrument drifts between calibrations.
- Consequence of an error: measuring a critical process variable is not the same as measuring a supporting one.
- Conditions of use: impacts, temperature, humidity, dirt, workload.
- Margin: the distance between the required tolerance and the uncertainty of the instrument.
- External requirements: regulations or customers that set a maximum.
How it is adjusted with data
Each calibration records the error found before any adjustment. With those records you can see the actual drift. If the instrument always turns up well within its limit, it is reasonable to consider extending the interval; if it turns up close to the limit or outside it, the interval should be shortened. There are formal analysis methods, such as those based on the proportion of instruments found in tolerance or on drift models, that give the decision a statistical basis.
What to record for the decision
- The indication error found before adjustment, not the one after.
- Whether an adjustment was made and which values changed.
- The conditions of use and relevant events, such as drops or repairs.
- The date of each calibration, to calculate drift per period.
When the routine does not apply
After an impact, a repair, an overload or a doubtful result, the instrument must be verified or calibrated even if the interval has not expired. The calendar is a guide, not a guarantee that the instrument will remain sound until that date.