What does “pretrigger” mean?

Pretrigger is the part of the capture that corresponds to the moments before the trigger event, and it is possible because a digital oscilloscope records continuously into a circular memory and keeps what happened beforehand.

What it is

On a digital oscilloscope, the trigger defines the time origin on the screen, but it does not have to be at the start of the capture. Pretrigger is the time interval before the trigger that the instrument includes in the waveform. Thanks to it, you can see what happened just before an event, which is often where the cause lies.

How it is possible

The instrument continuously digitizes and writes samples to a memory that works like a ring: when it fills up, new samples overwrite the oldest ones. When the trigger occurs, it keeps capturing for the time defined for post-trigger and then stops. What remains in memory includes samples from before the event. An analog oscilloscope cannot do this, because it draws the signal in the moment and does not store it.

How it is adjusted

The horizontal trigger position moves across the screen. If you place it toward the right, you see more time before the event; if you place it toward the left, more time after it. The total amount of time recorded depends on the record length and the time base: pretrigger and post-trigger share that memory. With a slow time base you can see a long time before the event, and with a fast one, only a short interval.

What it is for

  • Seeing which signal changed before a failure, such as a supply drop preceding a reset.
  • Measuring the starting level of a transition.
  • Studying what caused a spurious pulse by triggering on the pulse itself.
  • Capturing the beginning of an event in which the trigger occurs somewhat after its start, for example when the trigger level is halfway up an edge.

A practical detail

After the trigger is armed, the instrument must first fill the pretrigger portion before accepting an event. If the event occurs sooner, the trigger is delayed or lost. With a longer pretrigger, this waiting time also grows.