Why does mains hum appear on the trace?

Mains hum appears on the trace because the probe, its ground lead or the circuit itself picks up the installation's field at the line frequency, or because a ground loop drives mains current through the measurement connections.

A signal with a recognizable frequency

Hum is a component at the line frequency, which in Mexico is sixty hertz, often with its harmonics. It appears as a slow ripple on the signal. It is significant because it immediately indicates its origin: something in the installation is coupling into the measurement. Frequency analysis helps confirm that the component matches the line frequency.

How it gets in

  • Probe pickup: on an unconnected probe, or on a high-impedance input, the cable and tip act as an antenna for the electric fields of the mains. A probe with its tip in the air almost always shows hum.
  • Ground lead loop: the probe’s ground lead and the circuit form a loop in which a varying magnetic field induces a voltage.
  • Ground loop: if the circuit and the instrument are grounded through different paths, the potential difference between them drives mains current through the probe shield, and the resulting drop appears as signal.
  • Mains-connected supplies: noise from a switching power supply or a transformer gets in through the common mode.

How to locate the source

  1. Connect the tip to its own ground. If the hum persists, it is entering through the cable or the instrument and not through the circuit.
  2. Move the probe cable and watch whether the amplitude changes.
  3. Power the circuit from a battery, if possible, to see whether the hum comes from the mains supply.
  4. Try another outlet or put the instrument on another power circuit.

How to reduce it

Reduce the area of the ground lead loop with the shortest connection, measure with attenuating probes, use a differential probe if common mode is the problem, and avoid crossing measurement leads with power cables. Do not remove the instrument’s protective earth to suppress a ground loop: that creates an electric shock hazard. Measurements on energized circuits or at hazardous voltages must be made by qualified personnel, following the applicable safety regulations and using accessories with the appropriate category and rated voltage.