What causes an oscilloscope not to trigger stably?

Unstable triggering occurs because the instrument does not always find the same condition in the signal: the level is outside the signal, the source or coupling is not appropriate, noise causes multiple crossings, or the signal has several similar events per period.

What the trigger does

The oscilloscope starts each sweep when the signal meets a condition, usually a level crossing with a given slope. If the condition is met at a different point in the cycle on each sweep, the waveform appears to drift or flicker on the screen. The problem is always that the instrument is triggering at different points on the signal.

Most common causes

  • Level outside the signal: if the trigger level is above or below the values the signal reaches, there is no trigger; in auto mode, the instrument sweeps without synchronization.
  • Wrong source: the trigger is taking another channel, the external input or the line.
  • Coupling: a signal with a varying DC component or high-frequency noise may require AC coupling, high-frequency reject or noise reject.
  • Noise on the edge: if noise crosses the level several times on the slope, the trigger occurs at different points.
  • Complex signal: a patterned pulse train or bursts have many similar edges, and the instrument may pick a different one on each sweep.

How to stabilize it

  1. Set the mode to normal to see whether it is really triggering, and check that the level crosses the signal.
  2. Verify the trigger source and slope.
  3. Turn on noise reject or hysteresis, or limit the trigger bandwidth.
  4. Adjust the holdoff so that the instrument ignores the intermediate edges and triggers only at the start of each burst or pattern.
  5. For complex signals, use a more specific trigger: pulse width, pattern, protocol or a sequence condition.

What to check first

If there is no trigger at all, look at the level and the source. If there is a trigger but you see a double trace, it is usually noise or a pattern with repeated edges, and the solution is holdoff or a more selective trigger. If the trace drifts slowly, the signal and the trigger may come from sources of slightly different frequency.