The probe loads the circuit because it adds its resistance, its capacitance and the inductance of its ground connection in parallel; at low frequency the resistance is what counts, and at high frequency the capacitance dominates.

Measuring means disturbing

An instrument cannot observe without touching. To obtain the signal, the probe draws a small current from the node being measured. If the source impedance is low compared with that of the probe, this current is negligible. If the source has a high impedance, the presence of the probe changes what you want to measure.

The three components of loading

  • Input resistance: forms a divider with the source impedance and reduces the voltage at DC and low frequency. It matters at high-impedance nodes, such as a divider of large resistors or a sensor output.
  • Input capacitance: forms a low-pass filter with the source impedance. It slows edges, attenuates high-frequency components and can shift the frequency of an oscillator or destabilize an amplifier.
  • Inductance: that of the ground lead and the tip resonates with the input capacitance and adds ringing on fast signals.

What the engineer sees

Slower edges when the probe is connected, a circuit that stops oscillating or oscillates differently when touched, a DC level that drops, or a logic error that disappears when measured. This last clue is the most revealing: if the fault hides when you connect the probe, the probe’s capacitance is modifying the circuit’s behavior.

How to reduce it

  1. Use an attenuating probe, which increases the resistance and reduces the capacitance seen from the tip.
  2. For high frequencies, use a low-capacitance active probe.
  3. Measure at a low-impedance point whenever the circuit allows it, such as the output of an amplifier rather than the high-impedance node before it.
  4. Shorten the tip and the ground with the appropriate accessories.
  5. If the effect is still large, compare the measurement with a simulation or with a calculation of the divider formed by the source and the probe.

When it does not apply

For low-impedance nodes and moderate frequencies, the loading of an attenuating probe is negligible, and any differences you observe come from other causes, such as the ground connection or the bandwidth.