Why does a waveform show oscillations that do not exist?

A waveform shows nonexistent oscillations because of aliasing, interpolation across too few samples, resonance of the probe itself and its ground connection, or interference entering the measurement.

The four main causes

Aliasing

If the sampling rate is not at least twice the highest frequency in the signal, the samples cannot distinguish a high frequency from a low one. The result is a slow, stable oscillation that does not exist. It appears mainly when the time base is slow, because the instrument reduces the sampling rate to fill the memory. The trace changes when you change the time base, and that is a key clue: a real signal does not change frequency when you move the scale.

Interpolation artifacts

Between samples, the instrument draws a calculated curve. If an edge has few samples, interpolation can produce overshoot and ringing near the edge that the circuit does not have. This is an effect of the reconstruction, not of the signal.

Probe resonance

The inductance of the probe’s ground lead and the input capacitance form a resonant circuit. A fast edge excites it, and a damped oscillation appears immediately afterward. It changes when you shorten the ground connection.

Interference pickup

A loop formed by the probe and its ground picks up switching, radio or mains fields and adds them to the measurement. The oscillation appears even if the tip does not touch the circuit.

How to tell them apart

  1. Change the time base. If the apparent frequency changes, suspect aliasing.
  2. Increase the sampling rate or use peak detect mode.
  3. Shorten the probe ground or use a short-tip accessory. If the oscillation disappears or changes, it was the probe.
  4. Connect the tip to its own ground. If you still see the signal, it is pickup.
  5. Repeat with another probe or a different instrument.

When it is real

A real oscillation keeps its frequency when you change the time base and the probe, and it is explained by a resonant circuit in the design itself, such as a parasitic inductance and the capacitance of a switch. Before modifying the circuit, confirm that the phenomenon survives these changes.