The central idea
A real-time oscilloscope takes many consecutive samples in a single capture. For extremely fast signals, neither the converter nor the input chain can manage this. The sampling oscilloscope changes the approach: if the signal repeats identically in each period, it is enough to take a single sample per repetition, shifting the sampling instant little by little relative to the trigger. As many repetitions accumulate, the samples are placed in their relative positions and reconstruct a waveform that appears to have been captured with a much higher time resolution. This method is called equivalent-time sampling.
What makes it so fast
The bandwidth is set by the input sample-and-hold circuit, which only has to follow the signal for a very brief instant. The rest of the chain, which processes the samples, can work at a much lower speed. That is why it is used to characterize very high-speed communication signals, transmission lines with reflectometry, and microwave components.
Requirements and limitations
- Repetitive, stable signal: reconstruction assumes every repetition is the same. A non-repetitive signal or one with variations is reconstructed poorly.
- Trigger: an accessible trigger signal, a reference clock or a good-quality internal trigger is needed, because sampling is referenced to the trigger instant.
- It does not capture single events: a fault that occurs only once is lost.
- Acquisition time: gathering a complete waveform takes many repetitions, which is why it is slower to update.
- Trigger jitter: timing disorder between trigger and signal adds to the reconstructed waveform and limits the accuracy of time measurements.
An example
To see the eye diagram of a fast serial signal, a sampling oscilloscope takes samples across millions of bits and overlays them. The number of bits observed gives a very complete statistical picture of jitter and noise. A real-time oscilloscope can also produce eye diagrams if its bandwidth is high enough, but it does not always manage this with signals that are too fast.