How does a current probe work?

A current probe measures the magnetic field produced by the current in a conductor and delivers a proportional voltage, without interrupting the circuit, combining a Hall sensor for DC and low frequency with a transformer for the AC part.

Operating principle

A conductor carrying current creates a magnetic field proportional to it. The probe surrounds the conductor with a magnetic core. A Hall effect sensor located in the air gap detects the field, including the DC component. For higher frequencies, a winding acts as a current transformer. A feedback circuit combines the two stages so that the response covers everything from DC up to its bandwidth limit, and it delivers an output in volts per ampere or millivolts per ampere.

What to do before measuring

  1. Connect the probe to the channel and check that the instrument recognizes its scale factor, or configure it manually.
  2. With no conductor in the clamp, null the DC offset with the zero control.
  3. Degauss the core with the function provided. A large current or a previous overload leaves remanent magnetism that appears as a false offset.
  4. Place the clamp around a single conductor and observe the direction arrow.

Real limitations

  • Saturation: when the current, or more precisely the product of current and time, exceeds what the core can withstand, the output distorts and is no longer proportional.
  • Bandwidth and derating: at high frequencies, less peak or RMS current may be allowed than at low frequency.
  • Insertion: it adds a small series impedance to the conductor. This can be noticeable in very high-frequency, low-impedance circuits.
  • Noise and resolution: for small currents, the signal sits close to the probe’s noise.
  • Propagation time: it differs from that of voltage probes, which matters when calculating power.

How to interpret the reading

If the current measured with the clamp closed seems offset from the expected value, null the zero and degauss again. If a current peak appears flattened, suspect saturation. If you see noise synchronized with the switching even with no current, the field of a nearby conductor is influencing the probe. Measurements on energized circuits or circuits with hazardous voltages must be made by qualified people, following the applicable safety regulations and using accessories with the appropriate category and voltage rating.