How do you measure current in a conductor without opening the circuit?

To measure current without opening the circuit, a clamp-on current probe is used that measures the magnetic field around a single conductor, so the circuit remains intact and insertion is minimal.

How it is possible

Every current generates a magnetic field around the conductor. A current clamp surrounds the conductor with a magnetic core and measures that field. AC current clamps act as a current transformer. Those that also measure DC add a Hall effect sensor in the core’s air gap and usually combine both principles to cover everything from DC to high frequencies. The output is a voltage proportional to the current, which the oscilloscope displays in amperes if you configure the probe sensitivity.

Procedure

  1. Connect the probe to the oscilloscope and select the appropriate sensitivity for the magnitude of the current.
  2. With the clamp open and off the conductor, remove the offset: adjust the zero and degauss the core if the probe requires it.
  3. Open the clamp, place it around a single conductor and close it. Observe the direction arrow so that the sign is correct.
  4. Adjust the vertical scale and verify that the channel’s scale factor matches that of the probe.

What you must keep in mind

  • A single conductor: if you enclose phase and neutral, or go and return, the fields cancel and the reading is almost zero.
  • Saturation: a current above the core’s limit, or a large DC component on an AC-only clamp, distorts the measurement. Inrush peaks can saturate the core.
  • Bandwidth and range: these depend on the probe. A very small current sits close to the noise level; this is improved by winding several turns of the conductor through the clamp and dividing the reading by the number of turns.
  • Conductor position: centering it in the window reduces the error, and the fields of nearby conductors can have an influence.

When it does not apply

If you need a very small DC current with high accuracy, or if the conductor is not accessible, a series sense resistor or a different probe is advisable. Measurements on energized circuits or circuits with hazardous voltages must be made by qualified people, following the applicable safety regulations and using accessories with the appropriate category and voltage rating.