The definition and what it shows
Instantaneous power is p(t) = v(t) · i(t). It is not a constant value: it rises and falls with the shape of both signals. Its average over a whole number of cycles is the active power. In a reactive load, instantaneous power becomes negative during part of the cycle, which means the load is returning energy to the source. In a power switch, the product during the transition shows the switching losses.
Setup
- Measure the voltage with a probe suited to the level and to the circuit reference, and the current with a current probe or with a low-inductance series resistor of known value.
- Set the scale factor on each channel so that the instrument shows correct volts and amperes.
- Define a math function that multiplies the two channels. The resulting unit is the watt.
- Calculate the average of the resulting waveform to obtain the active power. Some instruments include a power analysis package that automates this.
Deskew
A voltage probe and a current probe do not have the same propagation delay. If the two signals are not aligned, the product is calculated with values from different instants. In a load with a phase angle, the resulting active power comes out wrong, and in a switching transition the loss area changes noticeably. The correction is made with the deskew function, by applying the same signal to both probes with a calibration fixture, or by adjusting according to the probe manufacturer’s data.
Other sources of error
- An unadjusted current probe zero, which shifts the entire product.
- Poor use of the vertical range, which reduces the resolution of both signals.
- Insufficient bandwidth in either of the two measurements.
- Averaging over a non-integer number of cycles.
How to interpret the result
If the calculated average does not match an independent measurement, first check the delay and the zero of the current probe. Measurements on energized circuits or circuits with hazardous voltages must be made by qualified people, following the applicable safety regulations and using accessories with the appropriate category and voltage rating.