How do you measure the noise and ripple of a power supply?

Power supply noise and ripple are measured with AC coupling, a minimal ground connection and a defined bandwidth, because the result depends more on how the probe is connected than on the instrument.

Ripple and noise are not the same

Ripple is the periodic component remaining at the output: at the rectified mains frequency in a linear supply or at the switching frequency in a switching supply. Noise also includes high-frequency spikes that appear at each switching event and other random components. A specification usually states both, with a declared measurement bandwidth, and without that figure the number is not comparable.

Measurement setup

  • AC coupling: it removes the DC level and allows the vertical sensitivity to be increased to see millivolts on an output of several volts. If the instrument has offset, DC coupling is an alternative that preserves the reference.
  • Bandwidth limiting: the channel filter separates ripple from very high-frequency noise. Record which limit you used.
  • Probe: a general-purpose attenuating probe is common, but it introduces relative noise when a small signal is measured; a direct coaxial connection or a unity-ratio probe can give a cleaner reading.

The ground connection decides the result

The long ground lead of the probe forms a loop that picks up the switching field and turns it into spikes that do not exist in the circuit. Remove the protective tip and use the ground spring or a short-tip adapter, touching the output terminal and its return as close as possible to the output capacitor. If you still see spikes when you rest the tip on its own ground, that is pickup in the measurement and not the supply.

How to read the trace

  1. Trigger on the switching or mains frequency so that the ripple remains stable.
  2. Measure the peak-to-peak value of the ripple with the filter on.
  3. Expand the time base on a switching edge to see the spikes and their duration.
  4. Repeat with the load at different levels, because ripple and noise depend on the current.

How to interpret the result

Ripple at the switching frequency usually points to the output filter and the series resistance of the capacitor. Spikes synchronized with the edges usually come from parasitic inductances. If the noise does not change when the supply is disconnected from the mains, the origin is external to it.