How do you set up a pulse width trigger?

A pulse width trigger makes the oscilloscope capture only when a pulse lasts less than, more than or exactly a defined time, which makes it possible to isolate defective pulses among many correct ones.

What this trigger does

An edge trigger reacts to a level crossing and does not distinguish whether the pulse is long or short. A pulse width trigger measures the time between the leading edge and the trailing edge of a pulse and compares that duration with a condition you define. Only if the condition is met does the oscilloscope trigger. In this way, a pulse that is too short or too long in a train of normal pulses becomes the event that anchors the screen.

Parameters you must set

  • Source: the channel carrying the pulse.
  • Polarity: positive or negative pulses. It defines which edge starts the width measurement.
  • Threshold: the voltage level used to decide when the pulse starts and ends. It should sit halfway up the edges, away from the noise.
  • Condition: less than, greater than, equal to, or inside/outside a range, with a reference time and, depending on the instrument, a tolerance.
  • Coupling and holdoff: holdoff prevents new triggers while the train is still in progress.

Procedure

  1. First observe the normal pulse with an edge trigger and measure its nominal width with cursors.
  2. Switch to pulse width trigger and set the polarity and threshold.
  3. Choose “less than” with a time slightly below the nominal to capture narrow pulses (glitches), or “greater than” with a longer time for stretched pulses.
  4. Use normal mode, not auto, so that the screen changes only when the event occurs.

What problems it helps you find

Spurious pulses on a clock or reset line, response times out of specification, anomalous PWM pulse widths and partial dropouts that do not reach the logic level. A pulse that does not cross the threshold (a runt) does not count as a pulse for this trigger, so in those cases a dedicated runt pulse trigger is used if the instrument offers one.

When it does not apply

If the signal is noisy and the threshold is crossed several times during an edge, the instrument will measure false widths. In that case, increase the hysteresis or noise rejection, or reduce the bandwidth with a filter to stop seeing multiple crossings.