Why the oscilloscope is a hazard
Most oscilloscopes connect the shield of their inputs to ground through the power cord. When the ground clip of a conventional probe is connected to a point in the circuit, that point becomes referenced to ground. If the point is at a different potential, a short circuit occurs through the instrument, with a risk of damage, arcing and injury.
Principles of a safe measurement
- Never defeat the protective ground of the oscilloscope, whether with adapters or by removing the ground pin. This practice leaves the enclosure and connectors exposed to hazardous voltage.
- Understand the circuit: know which points are referenced to the mains, which are floating, and what potential each one is at with respect to ground.
- Use probes with the appropriate voltage rating and measurement category, and accessories in good condition.
- Prefer differential probes or instruments with isolated channels when no point in the circuit can be connected to ground.
- Do not exceed the maximum voltage of the probe or the input, and take switching peaks into account, not just the RMS value.
What you see in practice
A differential probe provides an attenuated signal proportional to the difference between two points; bandwidth and common-mode rejection determine the quality in fast-switching circuits, where common-mode voltage swings are large. With the long ground lead of a passive probe, oscillations appear on the edges that do not exist in the circuit.
What to check first
- De-energize before connecting and wait for the capacitors to discharge.
- Check the condition of the leads, the probes and the oscilloscope’s ground.
- Connect with the circuit de-energized, with the tips secure and with hands away from live parts.
- Energize, measure, de-energize and disconnect.
Warning
Measurements on power or high-voltage circuits must be performed only by qualified personnel, following the applicable safety regulations and using the corresponding personal protective equipment.