What is aliasing and how is it avoided?

Aliasing is the error whereby an oscilloscope or data acquisition unit shows a fast signal as if it were a slower one, because the sample rate is insufficient to represent it. It is avoided by sampling faster or by limiting the frequency content of the signal.

What happens

Sampling means taking values at discrete instants. If the signal changes faster than the samples allow, several different frequencies produce exactly the same points, and the instrument cannot tell them apart. A component above half the sample rate is interpreted as a lower frequency: its alias. It is like the wheels of a vehicle that seem to turn slowly or backward in a film.

How to recognize it

  • A displayed frequency lower than expected, with no explanation in the circuit.
  • Waveforms that change appearance or frequency when the time base is changed, even though the signal has not changed.
  • An image that is unstable with the trigger and does not synchronize even though the level is correct.
  • Strange patterns on modulated or switching signals.

The most useful test is to vary the time scale: the real frequency does not change with it, whereas an alias does. Another clue comes from also measuring with a frequency counter or an independent reference.

How to avoid it

  1. Increase the sample rate: use a faster time base or a deeper memory that makes it possible to maintain the sample rate.
  2. Limit the bandwidth with the channel filter, if the high-frequency content is of no interest. Components that are removed can no longer turn into aliases.
  3. Use peak detection, which records maximums and minimums between samples and reveals fast events even when sampling is low.
  4. Filter before acquiring in data acquisition units, with an analog anti-aliasing filter, because once sampled, the alias cannot be separated from the rest.

Why it matters

Aliasing produces no warning: the screen looks normal. That is why it is advisable to know the effective sample rate at each setting and to distrust anything that does not match what you expect.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer