What is the trigger on an oscilloscope?

The trigger is the mechanism that decides the instant at which acquisition begins, so that every sweep starts at the same point on the signal; without a stable trigger, a periodic signal would appear shifted and blurred.

What it is for

An oscilloscope captures segments of time and draws them one on top of another. If each segment started at an arbitrary point on the signal, the image would move. The trigger defines a condition, a recognizable event in the signal, and each capture is aligned to it. The instant of the trigger event is the time reference for the screen.

The basic parameters

  • Source: the channel, external input or mains supply being monitored.
  • Level: the voltage the signal must cross.
  • Slope: whether the condition applies to a rising or a falling edge.
  • Coupling: DC, AC, or with high-frequency, low-frequency or noise rejection, so that the trigger circuit ignores irrelevant components.
  • Holdoff: the minimum time after a trigger during which new events are ignored. It is used for signals with bursts, where several similar edges would confuse synchronization.

Trigger modes

  • Auto: the instrument draws even without a trigger, which is useful for viewing a DC signal or for getting your bearings.
  • Normal: the display updates only when there is a valid event.
  • Single: it captures once and stops.

Advanced triggers

Besides edge, there are triggers on pulse width, short or runt pulses, time window, logic pattern or communication protocols. They make it possible to isolate specific events, such as a pulse narrower than it should be.

Pretrigger and position

Because the instrument stores data continuously, it can show what came before the event, which lets you see what happened before a fault.

What you see in practice

An image that slides indicates a level outside the signal, the wrong source, or noise causing false triggers. If the level is within the noise, triggering occurs at different points and jitter is visible; increasing the hysteresis or filtering stabilizes it. Remember that in normal mode, a frozen screen may mean that no event is arriving.