What do half digits mean on a multimeter?

A rating in digits and a half indicates display resolution: full digits can show zero to nine, and the half digit takes only limited values, normally zero or one. It does not indicate the instrument's accuracy.

How they are counted

A specification such as five and a half digits means the display has five positions that show any figure, plus one additional position, the most significant, that shows only a reduced value, for example zero or one. This partial digit extends the displayed range without adding another full position. On some instruments the partial digit can reach higher values, so it is advisable to check the manual for the maximum displayed.

What they indicate and what they do not

Digits describe resolution: the smallest change the display can represent. They do not describe accuracy, which is how close the reading is to the true value, or stability. An instrument can show many digits and still give a reading of low accuracy, because the last digits are dominated by noise, drift or calibration errors.

How to read the specification

Accuracy is stated as a combination of a term proportional to the reading and a fixed term associated with the range or a number of counts. To find out whether the extra digits are useful, compare the size of the uncertainty with that of the resolution. If the uncertainty spans several counts of the last digit, those digits provide no reliable information about the true value, although they can be useful for observing relative changes.

What you see in practice

  • The reading flickers in the last digit because of noise; with a longer integration time it stabilizes.
  • Available counts change with speed: a fast measurement usually offers fewer digits than a slow one.
  • An instrument with more digits makes it possible to distinguish small changes, which is useful in stability tests and comparisons.

An example

An instrument may be able to distinguish a very small change between two readings, but if its calibration has a larger error, that change does not correspond to a real difference relative to a standard. Resolution and accuracy answer different questions.