Do thermoelectric voltages cause errors in low-level measurements?

Yes. Wherever dissimilar metals join and there is a temperature difference, a small voltage appears through the Seebeck effect and adds to the signal, and in low-level measurements it can be comparable to or larger than what you want to measure.

The physical origin

When two conductors of different materials are joined and their ends are at different temperatures, a voltage proportional to that difference is generated. This is the Seebeck effect, the same one a thermocouple uses to measure temperature. In a measurement circuit, every junction of copper with another material, such as a solder joint, a binding post, a connector or a relay contact, can behave as a parasitic thermocouple.

Why it matters at low levels

These voltages are on the order of microvolts per degree for many combinations, so they do not matter when measuring volts, but they do when measuring microvolts, very low resistances or signals from sensors such as strain gauges and thermocouples. The parasitic voltage adds to the signal and is indistinguishable from it in a single reading.

What the engineer sees in practice

  • A nonzero reading with the inputs shorted.
  • An offset that changes when the connector is touched or an air current comes near.
  • A low resistance that reads different values depending on the polarity of the measurement current.
  • A slow drift that follows the room environment.

How it is reduced or compensated

  1. Use the same material along the entire path, with connectors and binding posts designed for low-level measurements.
  2. Maintain thermal equilibrium: protect the junctions from air currents and heat sources, and allow them to stabilize.
  3. Polarity reversal: measure with current in one direction and then the other, and average. The thermoelectric voltage does not change sign with the current, so it cancels out.
  4. Automatic zero compensation in the instrument, if available.
  5. As few and as symmetrical junctions as possible, so that their effects oppose each other.

How to interpret the result

If an offset remains when the source is reversed, it is thermoelectric or another constant offset; if it changes sign with the current, it is proportional to the current and forms part of the measurement. With mains or power signals, the work must be done by qualified personnel following the applicable regulations.