The mechanism
Every component has a temperature coefficient. When an instrument is switched on, its electronics dissipate power and the internal temperature rises until it reaches equilibrium with the ambient. While this happens, the voltage reference, which sets the accuracy of the entire instrument, changes slightly; the same happens with the resistor networks of the input divider, with amplifier offsets and with clocks.
Gradients, not just temperature
What matters is not only the absolute value but the differences between points. A thermal gradient along a junction of dissimilar metals produces thermoelectric voltages that add to the signal. In low-level measurements, this effect dominates, and it settles only when the interior reaches equilibrium.
Self-heating from the signal
The measurement itself produces heat. A precision resistor dissipating power changes value, hence the drift in a resistance measurement at high current or in high-voltage dividers. A temperature sensor powered with too much current heats up and reads above the actual temperature.
What the engineer sees in practice
- Readings that shift monotonically during the first minutes or hours after power-up, and then stabilize.
- Changes when a door opens, when the air conditioning changes or when a heat source comes near.
- Differences between morning and afternoon measurements with no other cause.
How to interpret the result
Manufacturers state a warm-up time before the specification is met, and a temperature coefficient for estimating the effect outside the calibration range. Drift within the specification is not a fault. Drift greater than expected may indicate a defect, an unstable environment or air currents.
How to reduce it
Switch the instrument on ahead of time, keep the temperature stable, avoid air currents over the connections, use cables and binding posts of the same material, and verify against a stable reference. Logging the ambient temperature along with the readings makes it easier to separate the thermal effect from other causes.