Why does a multimeter’s internal fuse blow?

A multimeter's fuse blows when more current flows through the current range than it can withstand, almost always because current was measured with the probes in the current terminals connected in parallel across a voltage source.

The main cause

On the current ranges, the multimeter presents a very low series resistance, because it is designed to be inserted into the circuit without altering it. If its current terminals are connected by mistake directly across a voltage source, as if it were a voltmeter, that near-zero resistance causes an enormous current and the fuse blows to protect the instrument. That is its job.

Typical situations

  • Leaving the lead in the current terminal after measuring current and moving on to measure voltage across a source: when changing measurements, the position of the lead is forgotten.
  • Measuring current in a circuit with high starting currents, such as motors or capacitive loads, which momentarily exceed the range value.
  • Selecting a current range that is too small for the circuit.
  • Measuring current in a high-voltage circuit, where the voltage across the fuse may exceed its interrupting capacity.

What you see in practice

After the fuse blows, the instrument appears to work normally on voltage and resistance, but the current measurement reads zero or appears as an open circuit. The resistance between the current terminals is very high, instead of a few ohms or less. This check, made with another multimeter and with the instrument out of circuit, identifies the blown fuse.

What to do correctly

  1. First identify and correct the cause, so as not to blow the replacement.
  2. Replace only with a fuse of the type specified by the manufacturer: rated current, rated voltage and breaking capacity. An ordinary fuse with a lower breaking capacity can be dangerous, and one with a higher current leaves the instrument unprotected.
  3. Never bypass a fuse.

How to prevent it

Always measure current with the circuit de-energized when inserting the instrument, check the position of the leads before each measurement and use a clamp meter whenever possible. On energized or power circuits, the work is the responsibility of qualified personnel following the applicable regulations.