How does a digital multimeter work?

A digital multimeter works by conditioning the input quantity into a DC voltage within the range of an analog-to-digital converter, which turns it into a number shown on the display.

The measurement chain

Every digital multimeter has a common structure. The input passes through conditioning that adapts it, a converter digitizes it, and a processor calculates and presents the result. What changes between functions is the conditioning ahead of the converter.

Conditioning by function

  • DC voltage: an input divider or an amplifier brings the signal into the converter’s interval. The input has a high impedance, so as not to load the circuit.
  • AC voltage: an RMS converter transforms the signal into a DC voltage proportional to its RMS value, either with analog circuits or by calculating it digitally from the samples.
  • Current: the current flows through an internal shunt resistor of known value and the voltage it produces is measured. That shunt introduces a small voltage drop into the circuit.
  • Resistance: the instrument injects a known current and measures the resulting voltage. With four wires it separates the current leads from the voltage leads to eliminate the wiring resistance.
  • Diodes and continuity: a small current is passed and the voltage drop is measured or compared with a threshold.

The converter

Higher-resolution multimeters use integrating converters, which average the signal over a period of time. If that time is a multiple of the power-line period, the line interference is rejected. Integration time is usually expressed in power-line cycles, and more cycles give more resolution and less noise, but a slower reading. Simple models favor speed and cost with lower-resolution converters.

Reference and accuracy

Accuracy depends largely on the stability of the internal voltage reference and of the precision resistors in the conditioning stage, which vary with temperature and time. That is why the instrument is calibrated periodically. Specifications are expressed as a percentage of reading plus a percentage of range, and they depend on the function.

What the engineer sees

Resolution in digits does not equal accuracy: an instrument with many digits can have poor accuracy. You must also consider the input impedance, the AC bandwidth and the measurement category, which indicates the type of circuit in which it is safe to use. Measurements on energized circuits are performed by qualified personnel with instruments of the appropriate category.