How do you measure a very small current?

A very small current is measured with a feedback picoammeter or electrometer, or with a very high-resolution multimeter on its lowest range, taking care with leakage, noise and the voltage drop the instrument introduces.

The challenge

A conventional ammeter measures the voltage drop across a small series resistance called a shunt. With tiny currents that drop would be minuscule and would be hidden by noise and offset. Increasing the resistance gives more signal, but it causes an appreciable drop in the measured circuit, known as burden voltage, which alters the current being measured.

How it is measured in practice

  • Picoammeter or electrometer: uses a feedback amplifier that holds the input almost at zero potential (virtual ground), so the burden voltage is minimal and the effective resistance seen by the source is very low, while the gain comes from a large feedback resistor.
  • High-resolution multimeter: can reach low current ranges, but its shunt generates a higher burden voltage.
  • Drop across a large resistor: the voltage is measured with a very high input impedance instrument; this is simple, although the loading introduced must be taken into account.

Sources of error

  • Leakage currents through dirty or damp insulators, connectors and circuit boards. They are reduced with guarding, low-leakage materials and cleanliness.
  • Triboelectric effect: friction or flexing of cables generates charges that are read as current. Special low-noise cables are used and are mechanically secured.
  • Electrostatic interference: charged objects or people moving nearby alter the reading; a properly connected metal shield prevents this.
  • Input bias current and thermal drift of the instrument, which set a minimum measurement level.
  • Source capacitance: increases noise and lengthens the response time.

How to interpret the result

With the input open and shielded, the reading should stay close to zero; that is the floor of the setup. A reading that changes when the wiring is touched or moved reveals electrostatic or triboelectric sources. A reading that takes a long time to stabilize indicates long time constants due to the circuit’s capacitance and resistance.

Common mistakes

  • Using ordinary unshielded cables.
  • Not waiting for settling.
  • Ignoring ambient humidity.
  • Measuring without knowing the instrument’s burden voltage.