How do you measure a very low resistance?

To measure a very low resistance, use a four-wire connection, a test current that is sufficient but controlled, and compensation of thermoelectric voltages, because at these values the wiring and contacts weigh as much as the component.

Why it is difficult

A very low resistance produces tiny voltage drops at moderate currents, and those drops compete with other effects of the same order: the resistance of leads and contacts, the voltages generated by junctions of dissimilar metals, and noise. With a two-wire method, the measurement is dominated by the wiring.

Method

  1. Four wires: one pair injects the current and another pair measures the voltage directly at the component, so the lead resistance does not enter the result.
  2. Clean, firm contacts: Kelvin clips or probes that press well, on surfaces free of oxide and paint. The sense point must be between the current contact and the component.
  3. Suitable test current: more current gives a higher voltage and a better signal-to-noise ratio, but it heats the component and changes its resistance. Choose the lowest current that provides a stable signal.
  4. Offset compensation: measure with the current in one direction and in the opposite direction and combine the results, so that the thermoelectric voltages, which do not change with the current direction, cancel out.
  5. Thermal stability: wait for the component and connections to reach equilibrium, and avoid air currents or contact with your hands.

Instruments

Micro-ohmmeters and high-resolution multimeters with a four-wire function and offset compensation are designed for this. With inductive components, such as windings, the current takes time to stabilize and you have to wait for the reading.

How to interpret the result

A reading that drifts slowly points to heating or thermal gradients. A reading that differs when the current is reversed reveals a thermoelectric offset. A reading that changes when the leads are moved indicates contact problems. Repeating the measurement with slightly different connection points and checking that the result holds gives confidence.

Common mistakes

  • Using long, unshielded sense leads in a noisy environment.
  • Raising the current until the sample heats up.
  • Measuring on a component that is still energized: the measurement must be made with the circuit de-energized and discharged.
  • Connecting the sense points before the current points.