How do you choose the right measurement range?

The right range is the smallest one that contains the signal without saturating, because an instrument's resolution and accuracy are specified relative to the range and improve when most of it is used.

Principle

An instrument converts the signal into a number with a converter of finite resolution. Each range spreads that resolution over the interval it covers. In a wide range, each converter step represents a larger quantity; in a narrow one, a smaller quantity. In addition, accuracy is usually specified as a percentage of reading plus a percentage of range, so using a range much larger than the signal makes the second term dominate the error.

How to choose it

  1. Know the order of magnitude of the signal. If it is unknown, start with the highest range and work down.
  2. Choose the lowest range that does not overflow, leaving a margin for peaks and variations. A signal at the edge of the range can saturate with a small fluctuation.
  3. Consider the maximum value, not just the average. In AC the peak matters, and the ratio between the peak and RMS values of the waveform (crest factor) can exceed the range limit.
  4. Check the protection level of the input: the maximum measurable value is not the same as the maximum the instrument can withstand without damage.

Autoranging and manual range

Autoranging selects the range by itself and is convenient for exploration. Its drawback is that it changes range during the measurement and takes longer; with signals close to the boundary between ranges it can hunt back and forth. For repeatable, automated or fast measurements, it is better to set the range manually.

What the engineer sees

A range that is too high produces readings with few significant figures and greater uncertainty. A range that is too low gives an overload indication. On oscilloscopes, the vertical scale is equivalent to the range: a signal that occupies few divisions uses few converter levels, and one that goes off screen is clipped.

How to interpret the result

If the same signal is measured on two ranges and the results differ by more than the expected uncertainty, check the zero, the instrument loading and the stability of the signal. Choosing a range is no substitute for consulting the instrument’s full specification.