General scope
IEC 61000-4-3 is a basic standard in the IEC 61000 series on electromagnetic compatibility. It establishes a test method for evaluating the immunity of equipment to radiated electromagnetic energy. Its purpose is to provide a reproducible procedure, not to define the severity levels a product must withstand: those levels and the acceptance criteria are determined by the applicable product or product-family standard.
What it describes
- Field source: use of antennas with an amplifier and generator, with amplitude modulation to simulate the effect of real transmitters.
- Test facility: anechoic or semi-anechoic chambers, with absorbers that ensure a controlled field.
- Field calibration: verification that a uniform field area exists where the equipment will be placed, checked with a probe without the equipment present.
- Equipment arrangement: position, cables and distance to the antenna, as well as the faces of the equipment that are exposed.
- Procedure: frequency sweep, sufficient dwell time, both polarizations and recording of behavior.
- Evaluation: classification of the equipment’s behavior according to the criteria defined in the product standard.
How it relates to other standards
Basic standards such as this one define the method. Product or product-family standards select the levels, frequency ranges and performance criteria that apply to each type of equipment. To find out how much a product must withstand, first identify which product standard applies to it.
What the engineer sees
The test reveals specific sensitivities: a frequency, a face of the equipment, a polarization or a cable. With this information the entry path is located and corrected, for example with filters at a sensor input or with shielding improvements.
When to consult the current edition
The standard has undergone revisions that change how the field is calibrated, how frequency ranges are treated and which techniques are permitted. When working on a specific project, consult the current edition and the one referenced by the product standard that applies to your market, and do not rely on general descriptions such as this one for procedural details.