Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

What is an absorbing clamp used for?

It is used to measure the disturbance power that equipment sends along its power cord or other cables at frequencies where measuring the radiated field is inconvenient. It is slid along the cable until the maximum is found, and it replaces the antenna as the evaluation method.

The idea behind the method

At high frequencies, small equipment such as household appliances or tools emits mainly through its cables, which act as antennas. Instead of measuring the field at a distance, one can measure the radio-frequency power that the equipment delivers to its cable. This quantity is called disturbance power, and it is related to what the cable would radiate.

What an absorbing clamp is

It is a set of ferrite rings surrounding the cable, incorporating a current transformer and a set of absorbers. It performs three functions:

  • it measures the radio-frequency current on the cable;
  • it presents a defined impedance, so that the measured current converts to power;
  • it absorbs the energy traveling toward the opposite side, isolating the setup from whatever happens beyond the clamp.

How it is used

  1. The equipment is placed on a non-metallic support at the height stated in the standard, with the cable stretched out in a straight line.
  2. The clamp is placed around the cable and connected to a receiver.
  3. For each frequency, the clamp is moved along the cable, from the equipment outward, until the maximum is found.
  4. That maximum is recorded as the disturbance power and compared with the limit.

Why it is slid along the cable

The current on the cable forms a standing wave. Its value depends on position, and the maximum changes with frequency, so a fixed position is not enough. The clamp seeks the maximum with a travel that covers the necessary length.

When it is used

It appears in product standards for household appliances, tools and similar equipment, in the frequency range where the cable dominates the emission. It is not suitable for equipment whose radiation comes mainly from the enclosure or from many cables. The method requires the equipment to remain in a stable operating condition during the sweep and the cables to be in the defined arrangement.

When measuring on energized power cables, the work must be carried out by qualified personnel, in accordance with safety regulations.