A reverberation chamber is a closed metal cavity with a mode stirrer that changes the field distribution; averaging over the stirrer positions yields a statistically uniform, isotropic field, which is used to test immunity or to measure total radiated power.

A cavity used on purpose

The walls of a reverberation chamber are metal and reflect almost all of the energy. Instead of avoiding resonances, the design exploits them: the cavity supports many resonant modes, and their number grows with frequency. Above a minimum frequency, which depends on the size of the chamber, there are enough modes to describe the field statistically.

The mode stirrer

A set of rotating metal paddles (the mode stirrer, or simply stirrer) changes the boundary conditions. With each position, the spatial distribution of the field changes. Measuring over a set of positions and averaging yields a field whose mean value is the same at any point in the working volume and for any orientation or polarization.

Uses

  • Immunity: an antenna transmits, and a field strength measured with the prior calibration is maintained. High fields are achieved with moderate amplifier power because the energy stays confined.
  • Emissions: the equipment operates inside the chamber and the total radiated power is measured, regardless of the direction in which it radiates, unlike a site with an antenna at a distance, where the directional maximum is sought.
  • Shielding: the chamber makes it possible to evaluate the effectiveness of enclosures and gaskets with a field that attacks from all directions.

How it differs from an anechoic chamber

In an anechoic chamber the walls absorb, and the wave arrives from a known direction with a defined polarization. In a reverberation chamber there is no preferred direction: the equipment is exposed to all of them. That is why the radiation pattern cannot be obtained, nor can the point of emission be identified, but the total can.

Limitations

  • It does not work well below the lowest usable frequency.
  • Loading (equipment, absorbers, cables) reduces the quality factor and must be taken into account in the calibration.
  • Enough stirrer positions must be used, and they must be statistically independent.

Each product standard states whether it accepts the reverberation chamber method as equivalent to others.