A cavity used on purpose
The walls of a reverberation chamber are metal and reflect almost all of the energy. Instead of avoiding resonances, the design exploits them: the cavity supports many resonant modes, and their number grows with frequency. Above a minimum frequency, which depends on the size of the chamber, there are enough modes to describe the field statistically.
The mode stirrer
A set of rotating metal paddles (the mode stirrer, or simply stirrer) changes the boundary conditions. With each position, the spatial distribution of the field changes. Measuring over a set of positions and averaging yields a field whose mean value is the same at any point in the working volume and for any orientation or polarization.
Uses
- Immunity: an antenna transmits, and a field strength measured with the prior calibration is maintained. High fields are achieved with moderate amplifier power because the energy stays confined.
- Emissions: the equipment operates inside the chamber and the total radiated power is measured, regardless of the direction in which it radiates, unlike a site with an antenna at a distance, where the directional maximum is sought.
- Shielding: the chamber makes it possible to evaluate the effectiveness of enclosures and gaskets with a field that attacks from all directions.
How it differs from an anechoic chamber
In an anechoic chamber the walls absorb, and the wave arrives from a known direction with a defined polarization. In a reverberation chamber there is no preferred direction: the equipment is exposed to all of them. That is why the radiation pattern cannot be obtained, nor can the point of emission be identified, but the total can.
Limitations
- It does not work well below the lowest usable frequency.
- Loading (equipment, absorbers, cables) reduces the quality factor and must be taken into account in the calibration.
- Enough stirrer positions must be used, and they must be statistically independent.
Each product standard states whether it accepts the reverberation chamber method as equivalent to others.