The problem it solves
The impedance of the electrical network depends on the location, the time of day and what is connected, and it also carries its own noise. If the disturbance from equipment were measured directly at the outlet, the result would depend on those conditions. The LISN (also called an artificial mains network) establishes a repeatable measurement environment.
What it does
- It passes the mains frequency and load current to the equipment with low loss.
- It presents to the equipment, in the radio-frequency range, a nominal impedance defined by the standard.
- It blocks noise coming from the network, with internal filters, so that it is not measured as if it came from the equipment.
- It extracts the disturbance from the equipment with a coupling capacitor and delivers it to a measurement port of known impedance.
Structure
Each conductor has a series inductor toward the network, an isolating capacitor to ground and a branch to the measurement port through a coupling capacitor. The port is connected to the receiver with a coaxial cable. There are versions for each conductor separately or with a switch between line and neutral. Three-phase networks follow the same principle.
How it is used
- The LISN case is bonded to the reference ground plane.
- The equipment is connected to the load side and the network to the supply side.
- The measurement port is connected to the receiver, and the other ports not being measured are terminated with the corresponding impedance.
- At the receiver, the network factor correction is applied, which includes the attenuation of the coupling and of the cables.
What to check first if the results look odd
- An open or poorly terminated port alters the impedance and distorts the reading.
- A deficient ground connection changes the coupling to the plane.
- If the correction factor is missing, the level appears lower or higher than the real value.
The LISN operates connected to the mains and can have leakage currents to ground. Its installation and use must be carried out by qualified personnel in accordance with the applicable regulations.