How is conducted disturbance measured on a power network?

The equipment is connected to the network through a line impedance stabilization network (LISN), which presents a defined impedance and delivers to the receiver the disturbance voltage of each conductor with respect to ground, using the detector and frequency range set by the applicable standard.

What is measured

Conducted disturbance is the radio-frequency voltage that equipment injects into its power cable. It is not measured directly on the public network because that network’s impedance varies with location and time of day and because it contains its own noise. The LISN solves both problems: it presents a stable impedance to the equipment and isolates the setup from the network.

Setup

  • The equipment is placed on a table or on the floor, at a defined distance from a reference ground plane and from the LISN.
  • The power cable is arranged according to the standard, folding the excess instead of coiling it.
  • The LISN is bonded to the ground plane and its measurement port is connected to the receiver by a coaxial cable.
  • Unused measurement ports are terminated with the receiver impedance.

Procedure

  1. With the equipment off, measure the ambient environment to establish the noise floor and external signals.
  2. Switch on the equipment in its highest-emission mode.
  3. Sweep the frequency range with a peak detector to identify the critical points.
  4. Repeat at the points close to the limit with the detectors required by the standard: quasi-peak and average.
  5. Carry out the procedure for each line conductor and for the neutral.
  6. Add the correction factors of the cable, attenuator and LISN to obtain the true value.

How to interpret the result

The result is compared with the limit in the product standard, in decibels referred to one microvolt. An emission that exceeds the limit with the peak detector may comply with quasi-peak and average, so the comparison must be made with the corresponding detector. Isolated peaks at multiples of the power supply switching frequency indicate that the input filter is insufficient. A high level on a single phase suggests asymmetry.

The measurement is made with mains voltage present, so it must be carried out by qualified personnel, in accordance with the applicable safety regulations.