Origin and what it represents
When a contact opens an inductive load (relay, contactor, solenoid, motor), the current cannot stop immediately. The contact bounces and many small successive arcs occur, each with a very fast rising edge. The result is groups of pulses that propagate along nearby power wiring. The fast burst test (EFT/B, electrical fast transient/burst) reproduces this phenomenon in a controlled way.
What the test signal looks like
- The pulses have nanosecond rise times and broad spectral content.
- They are grouped into packets with a defined repetition frequency and a pause between packets.
- The energy of each pulse is low; what makes it aggressive is the speed and repetition, not the energy.
How it is applied
At power ports, the pulses are injected through a coupling and decoupling network that prevents them from reaching the laboratory mains. On signal and control lines, a capacitive coupling clamp is used that surrounds the cable without touching it. The equipment is placed on a reference plane with the cabling arranged according to the applicable standard. Generator verification uses attenuators and an oscilloscope with sufficient bandwidth.
What the engineer sees in practice
Normally there is no damage. Typical symptoms are microcontroller resets, communication errors, erroneous sensor readings, flickering displays or digital inputs that change state. The cause is usually capacitive and inductive coupling of the pulses into sensitive traces, or a disturbance of the ground reference that the circuit interprets as a signal.
How to interpret the result
The result is classified according to the performance of the equipment during and after the test: normal operation, degradation that recovers on its own, degradation that requires operator intervention, or loss of function. The acceptable criterion is defined by the product standard, not the test standard.
The usual corrections are filtering at the input port, good management of ground references, reduction of loop length and area, and protection with ferrites or suppressors on signal lines. Tests at mains voltages must be carried out by qualified personnel, following the applicable regulations.