What linear region means
An amplifier is linear as long as the output is proportional to the input: the gain does not depend on level. As the input increases, the device approaches its limit, gain decreases and harmonics and intermodulation products are generated. Characterizing the linear region means measuring how far the level can be raised before that happens and how well the amplifier behaves until then.
Small-signal measurements
With a network analyzer and a power level low enough not to cause compression, the S-parameters of the amplifier are measured: gain is the forward transmission parameter, and input and output match are the reflection parameters. Gain flatness across the band is also observed and, if necessary, phase. The bias point and temperature must be controlled, because gain depends on both.
Power sweep and compression
A single-tone signal at the desired frequency is applied and its level is increased gradually while the output power is recorded. The output-versus-input curve is a straight line in the linear region and flattens near saturation. The one-decibel compression point is the level at which actual gain is one decibel below linear gain, and it is used as a practical reference for the end of the linear region. The same measurement can record phase variation with level, which matters in systems with phase modulation.
Intermodulation
With two closely spaced tones, a nonlinear amplifier generates intermodulation products at adjacent frequencies. Their ratio to the main tones indicates linearity. The third-order intercept point summarizes this information in a single extrapolated number. Other linearity analyses use modulated signals.
What to watch in the setup
- Place attenuators at the output and calibrated couplers so that the meter or analyzer does not receive more power than its limit.
- Use a filter between the source and the amplifier if the source has harmonics, because they would be confused with those generated by the amplifier.
- Measure in pulsed mode when heating can alter the result, and wait for the temperature to stabilize.
- Calibrate the level at the amplifier plane, not at the source.
Power amplifiers handle levels that can cause burns and dangerous exposure. The measurement must be performed by qualified personnel following the applicable regulations.