How Does a Swept Spectrum Analyzer Work?

It is a superheterodyne receiver that mixes the signal with a swept local oscillator, filters it with a narrow intermediate-frequency filter and detects it, displaying the level of each frequency as the sweep advances.

The signal chain

  1. Input attenuator. Adjusts the level to protect the mixer and avoid distortion.
  2. Input filter or preselector. Limits the band and reduces unwanted responses, such as the image frequency.
  3. Mixer and swept local oscillator. The local oscillator varies its frequency along a ramp. Mixing translates the part of the input spectrum that corresponds to each instant to a fixed intermediate frequency.
  4. Intermediate-frequency filter. This is the resolution filter: it passes only what falls around it. Its width determines how two nearby signals appear.
  5. Envelope detector. Converts the filtered signal into a level.
  6. Video filter. Averages the detected level and smooths the noise on the trace.
  7. Display. The ramp that sweeps the local oscillator also drives the horizontal axis, so each position corresponds to a frequency.

What the principle implies

  • At each instant the instrument looks at only a narrow band. A brief event that occurs outside the band being observed at that moment is not detected. Intermittent signals may not appear, or may appear at varying levels from sweep to sweep.
  • The sweep cannot be arbitrarily fast: the narrow filter needs time to settle, which ties the resolution bandwidth to the sweep time.
  • The on-screen level is relative to the instrument’s calibration, and its level accuracy is not that of a power meter.

Measurement limitations

The mixer generates its own distortion products if the input signal is very strong; to check whether distortion is real, add attenuation and see whether it changes. The instrument’s noise floor sets the smallest visible signal, and it drops with a narrower filter.

When another type of analyzer is the better choice

For transient events or frequency-hopping signals, an analyzer based on the fast Fourier transform or a real-time analyzer captures the entire interval instead of a sequential sweep.