How does a Rogowski coil work?

A Rogowski coil is a winding with no magnetic core placed around the conductor. By Faraday's law, its output voltage is proportional to the derivative of the current, and an integrator converts it into a signal proportional to the current.

Physical basis

A current-carrying conductor creates a magnetic field around it. If the field passes through a coil, the voltage induced in the coil is proportional to the change in flux, according to Faraday’s law. A Rogowski coil is wound uniformly on a non-magnetic former, with a return conductor that cancels the flux that would otherwise pass through the coil’s single loop. By Ampère’s law, the output voltage depends only on the enclosed current and on a mutual inductance constant: v = M · di/dt.

Why an integrator is needed

The output is proportional to the derivative of the current, not to the current. To recover the waveform, the signal is integrated with an analog or digital electronic integrator. The output grows at higher frequencies, so the sensor is very sensitive to high-frequency components; the integrator must behave well at low frequency so that it introduces neither drift nor phase error.

Advantages of having no core

  • No saturation: the response is linear even when the current is very large.
  • Wide dynamic range: the same coil measures from small currents up to very large ones, with the help of the integrator’s range.
  • Flexibility and low weight, so it can be placed on conductors of varied shapes.
  • Wide bandwidth, useful for current harmonics and transients.

Limitations

It does not measure direct current, because with no change in flux there is no induced voltage. Its accuracy depends on the uniformity of the winding, and it is sensitive to conductor position and to the fields of nearby conductors. The output signal is small, so cabling and shielding matter. Sensitivity and accuracy specifications are specific to each manufacturer.

How to interpret the result

The measurement is only as good as the coil and integrator as a set, which are usually calibrated together. If either one is changed, the scale factor must be verified. Practical use on busbars is described in another guide. Installation on energized conductors must be done by qualified personnel.