Three-Cup Insulating Oil Dielectric Strength Tester

Equipment for testing the dielectric strength value of various insulating oils, with three test cups. For laboratories and transformer maintenance personnel.

Description

The three-cup insulating oil dielectric strength tester is used to test the dielectric strength value of various insulating oils. It complies with the ASTM D1816 and ASTM D877 standards. It can test six types of oil samples at the same time, which is convenient and fast, with accurate results. You can choose from three types of electrode heads (flat, spherical and hemispherical) according to your needs.

Functions and features

  1. The instrument is designed for three cups.
  2. 5.4-inch LCD screen, with Chinese-language menu for human-machine dialogue, temperature display function, humidity measurement and clock.
  3. Operation is simple, and it automatically completes the raising, holding, stirring, static setting, calculation, data storage and printing operations.
  4. It has a power-loss storage function, and test results are saved automatically.
  5. It has overvoltage, overcurrent, limit and grounding alarm functions to protect the operator’s personal safety.
  6. The unique waveform setting function eliminates harmonic interference for accurate measurement.
  7. It is controlled by a dual CPU and a PLC safety instrument, stable and reliable.
  8. It has a 232 interface and a Bluetooth function for transmitting data to the computer.

Main technical parameters

Output voltage 0 ~ 80 kV (0-100 KV can be configured)
THDu <1%
Voltage rise rate 2.0~3.5 kV/s (Δ = 0.5 kV/s)
Step-up capacity 1.5 kVA
Measurement accuracy ±2%
Supply voltage 220 V AC ±10%
Power frequency 50Hz ±2%
Power 200w
Applicable temperature 0~45℃
Applicable humidity <85% RH
Width x height x depth 410×390×375 (mm)
Net weight ~ 32 kg

Operation

The three-cup transformer oil dielectric strength test set has the advantage of simple operation: the operator only needs a simple setup, and the instrument configures itself automatically according to the voltage test of oil samples 1-3. Each oil sample, each breakdown voltage value and the number of cycles are stored automatically. When the test ends, the thermal printer can print the breakdown voltage value and the average value.

Precautions

  1. Before the test, the selection and placement of the oil samples and the distance between the electrodes must comply with national standards and industry standards.
  2. After the power is connected, the operator and all other personnel are strictly prohibited from touching the housing, to avoid hazards.
  3. If any abnormality is found while using the instrument, the power supply must be cut off immediately.
  4. A new or freshly cleaned oil cup undergoes 24 breakdowns before the test. When no test is being performed, the oil cup is kept immersed in clean oil.
We do real tests, on video! Watch them on our YouTube channel →